1.
SPECIMEN:
- Blood sample
2. EQUIPMENT & REAGENTS
- Equipment:
- Malarial parasite kit (also includes reagent).
3. PROCEDURE:
- Apply blood from the tube into the center of the cassette sample well by pressing the end of the tube onto the absorbent paper in the well.
- Allow up to 30 seconds for the specimen to be absorbed before adding the reagent.
- Add 5 drops (about 8 µl) of the reagent into the sample well.
- Results are read in 5 – 10 minutes for strong positives.
- Wait up to 15 minutes for weaker positives and to make sure negatives are confirmed.
- Do not read results after 15 minutes.
- If the membrane does not clear sufficiently after 2 minutes, one more drop of reagent may be added to the (sample) well of the cassette.
- Observe the result:
- The test is negative if only the Control line appears.
- If neither control or test line appears, the result is invalid.
- The test is positive when both control and test lines appear in the test window.
- Prepare lab report of the observations.
No comments:
Post a Comment